Specifications
Diamondback360®CoronaryOrbitalAtherectomySystem
26
4. Lock the VIPERWIRE guide wire in place by pressing down on the
guide wire brake lever as the crown will not spin if the guide wire
brake is unlocked.
5. Verify that saline is still flowing freely out of the saline sheath tip.
Verify that the saline tubing is properly connected to the saline bag,
that the saline tubing routes correctly through the saline tubing
guides, and that the saline tubing is properly connected to the OAD.
6. Press and release the On/Off button located on top of the crown
advancer knob to activate crown rotation. The OAD is preset to low
speed, and the illuminated LED on the OAD will indicate that the
OAD is operating at low speed.
7. Check that the flow of saline is increasing and that the shaft and
crown are beginning to rotate.
8. Immediately press and release the On/Off button to stop the shaft
and crown from rotating and to complete the test.