9.5. Preparing the bag of saline and VIPERSLIDE Lubricant
1. Ensure that the OAS pump is powered off by pressing the Master Power
switch on the back of the OAS pump to off and ensure that no LEDs are
illuminated on the OAS pump panel (Figure 4).
2. Prepare a full 1000 mL bag of normal saline solution with VIPERSLIDE
Lubricant. Refer to the VIPERSLIDE Lubricant Instructions for Use for
lubricant preparation instructions.
3. Hang the prepared saline bag with VIPERSLIDE Lubricant from the low saline
level sensor on the standard IV pole.
Caution: Do not use glass bottles for the saline solution with VIPERSLIDE
Lubricant or hang multiple saline bags from the low saline level sensor as this
will disable the Low Saline Information signal.
Figure 5. Placing the saline tubing within the OAS pump___________________
A. Saline tubing positioners
B. Saline tubing
C. V-guides
D. Pump door