Specifications
Diamondback360®CoronaryOrbitalAtherectomySystem
20
9. Equipment, set up, and test
9.1. Equipment
In addition to OAS components, equip the operating room with the following:
Guide catheter - see Table 1 for guide catheter size recommendations
Standard IV pole with five wheels and a 20 inch diameter base
1000 mL bag of sterile normal saline
Fluoroscopic imaging equipment
Standard 110 V hospital grade, electrical wall outlet
Other equipment, as needed, for interventional procedures
9.2. OAS Pump set up
1. Use the IV pole screw clamp to attach the OAS pump to a standard IV pole
making sure to attach the OAS pump to the IV pole at a distance not greater
than 60 in (153.0 cm) from the floor to the top edge of the OAS pump.
2. Hang the low saline level sensor and cord, by the closed loop, from the
horizontal arm of the standard IV pole.
3. Plug the low saline level sensor connector into the back of the OAS pump.
4. Verify that the power cord is connected to the back of the OAS pump.
5. Insert the other end of the power cord into the electrical wall outlet.
9.3. Preparing the VIPERWIRE Advance
Coronary Guide Wire
1. While using sterile techniques, open the VIPERWIRE guide wire packaging
pouch and remove the packaging tube.
2. Remove the VIPERWIRE guide wire and VIPERWIRE guide wire torquer
from the packaging tube as follows:
a. Locate the proximal VIPERWIRE guide wire retainer on the inside of the
packaging tube. This will expose the proximal end of the VIPERWIRE
guide wire.
b. Removing the VIPERWIRE guide wire – distal end first: advance the
proximal end of the VIPERWIRE guide wire into the packaging tube. This
will expose the distal end and spring tip of the VIPERWIRE guide wire.
Grasp the exposed distal end of the VIPERWIRE guide wire and gently
pull the VIPERWIRE guide wire out of the packaging tube. Use care to