Operational Considerations
3 Operational
Considerations
3.1 Limit Switches
M-116 stages are optionally available with two optical limit
sensors with TTL drivers. Limit switch outputs are active-high
for the DC-motor version. and active-low for the stepper motor
version. Models without the optional limit switches put the TTL
level indicating “no overtravel” on the limit switch lines. The
distance between the limit switches must be specified with the
order. If no reference switch is required, the limit switches may
be anywhere between 0º and 330º; if the reference is installed,
then the lower limit switch must be at ≥ 30º. Limit switch
installation is accurate to ±2°, limit switch repeatability is 0.004°.
3.2 Position Reference Signal Sensors
A position reference sensor is located approximately in the
middle of the operating range and can be used to reference the
absolute position of the stage within 0.004° accuracy. Always
approach the reference sensor from the same side to reach the
same position.
The reference sensor provides a static signal level which
depends on the platform position. If the platform position is less
than a half-turn counterclockwise from the reference switch
position, the reference signal is +5 V, while if the platform is
less than a half-turn clockwise from the reference switch, the
signal level is 0
V.
3.3 Recommended Motor Controllers
M-116.DG and M-116.DGH stages can be used with C-842, C-
843, C-844, C-848 and Mercury (C-860, C-862) controllers, the
M-116.2S and M-116.2SH with the C-600 and C-630 stepper-
motor controllers.