Specifications

D1600 Data Sheet B-3
Figure B-3. Controlling Hysteresis For Bipolar Signals.
R
Open
Ø
34 V
hysteresis
R (in K) =
0.5 - V
hysteresis
V
switching
= 2.5 -
The hysteresis control may also be connected to ground (GND), which
produces another set of switching levels. This connection is shown in figure
B-4. If the HYSTR terminal is shorted to GND the nominal switching point is
1.6V with ±5mV of hysteresis.
To measure AC signals super-imposed on a DC value, the input may be AC
coupled by simply placing a capacitor in series with the +IN terminal. The
Figure B-4. Controlling Switching Level and Hysteresis.
V
switching
± V
hysteresis
2.5V ± 0.5V
1.7V ± 5mV
For V
hysteresis
>5mV and <0.5V:
14
17 + R