Datasheet
Timing in HR Mode
The internal FIFO stores up to 32 samples, so that the system
processor does not need to read the data after every sample.
In HR mode (Figure 5), unlike in SpO
2
mode, temperature
information is not necessary to interpret the data. The user
can select either the red LED or the infrared LED channel for
heart rate measurements.
Figure 5. Timing for Data Acquisition and Communication When in HR Mode
Table 16. Events Sequence for Figure 5 in HR Mode
EVENT DESCRIPTION COMMENTS
1 Enter into Mode
I
2
C Write Command sets MODE[2:0] = 0x02. Mask the A_FULL_EN
Interrupt.
2 FIFO is Almost Full, Interrupt Generated Interrupt is generated when the FIFO has only one empty space left.
3 FIFO Data is Read, Interrupt Cleared
4 Next Sample is Stored
New sample is stored at the new read pointer location. Eectively, it is
now the rst sample in the FIFO.
INT
I
2
C Bus
LED
OUTPUTS
IR
~
~
~
SAMPLE #1 SAMPLE #2 SAMPLE #3 SAMPLE #30 SAMPLE #31
1 2
3 4
15ms TO 300ms
IR IR IR IR IR IR
www.maximintegrated.com
Maxim Integrated
│
27
MAX30102 High-Sensitivity Pulse Oximeter and
Heart-Rate Sensor for Wearable Health