Datasheet
Timing for Measurements and Data
Collection
Slot Timing in Multi-LED Modes
The MAX30102 can support two LED channels of sequential
processing (Red and IR). Table 14 below displays the four
possible channel slot times associated with each pulse width
setting. Figure 3 shows an example of channel slot timing for
a SpO
2
mode application with a 1kHz sample rate.
Table 14. Slot Timing
RED LED
660nm
Red On
69μs
Red Off
931μs
IR On
69μs
IR Off
931μs
358μs
INFRARED LED
880nm
Figure 3. Channel Slot Timing for the SpO
2
Mode with a 1kHz Sample Rate
PULSE-WIDTH SETTING (µs)
CHANNEL SLOT TIMING (TIMING
PERIOD BETWEEN PULSES) (µs)
CHANNEL-CHANNEL TIMING
(RISING EDGE-TO-RISING EDGE)
(µs)
69 358 427
118 407 525
215 505 720
411 696 1107
www.maximintegrated.com
Maxim Integrated
│
25
MAX30102 High-Sensitivity Pulse Oximeter and
Heart-Rate Sensor for Wearable Health