Datasheet

Multi-LED Mode Control Registers (0x11–0x12)
In multi-LED mode, each sample is split into up to four time slots, SLOT1 through SLOT4. These control registers determine
which LED is active in each time slot, making for a very flexible configuration.
Each slot generates a 3-byte output into the FIFO. One sample comprises all active slots, for example if SLOT1 and
SLOT2 are non-zero, then one sample is 2 x 3 = 6 bytes.
The slots should be enabled in order (i.e., SLOT1 should not be disabled if SLOT2 is enabled).
Table 9. Multi-LED Mode Control Registers
REGISTER B7 B6 B5 B4 B3 B2 B1 B0
REG
ADDR
POR
STATE
R/W
Multi-LED
Mode Control
Registers
SLOT2[2:0] SLOT1[2:0] 0x11 0x00 R/W
SLOT4[2:0] SLOT3[2:0] 0x12 0x00 R/W
SLOTx[2:0] Setting WHICH LED IS ACTIVE LED PULSE AMPLITUDE SETTING
000 None (time slot is disabled) N/A (O󰀨)
001 LED1 (Red) LED1_PA[7:0]
010 LED2 (IR) LED2_PA[7:0]
011 None N/A (O󰀨)
100 None N/A (O󰀨)
101 Reserved Reserved
110 Reserved Reserved
111 Reserved Reserved
www.maximintegrated.com
Maxim Integrated
21
MAX30102 High-Sensitivity Pulse Oximeter and
Heart-Rate Sensor for Wearable Health