Data Sheet
Conductivity Electrode User Manual
1. Application
This  conductivity  electrode  is  the  measuring  element  of  the  conductivity  meter.  It  used  to 
measure the conductivity value of the aqueous solution or as a conductance titration.
2. Model and Main Technical Parameters
Model
DJS-0.1
260
DJS-1
DJS-10
Vessel 
Constant
0.1±0.02
0.6±0.2
1±0.2
10±2
3. Use and Maintenance
1. There  are  two  kinds  of  conductivity  electrode,  shiny  electrode  and  platinum  black 
electrode. platinum-plated black aims to increase the effective area of the electrode sheet 
and relieve being polarized. So in the measurement of large conductivity solutions, using 
a platinum black electrode is more appropriate.
2. For  the  platinum  black  electrode,  before  use,  it  can  be  immersed  in  deionized  water  to 
prevent the platinum black inerting.
3. When  platinum  black  coating  off  or  fade,  then  you  must  re-plating  platinum  black  to 
ensure the accuracy of the measurement readings.
4. The cell constant marked on the electrode is just for reference. If you want to test the cell 
constant is accurate or not, you can  lookup the corresponding conductivity value K 
according to the concentration and temperature of the tested solution. Then  the cell 
constant Q is calculated based on the measured resistance R .
The formula is Q = K × R .
See table below for details:
Model
Solution 
Concentration
Solution 
Temperature( ℃)
Conductivity Value
25
1.468×10
-4
DJS-0.1
0.001mol/L KCL
30
1.64×10
-4
25
1.413×10
-3
260、DJS-1
0.01mol/L KCL
30
1.522×10
-3
25
1.28×10
-2
DJS-10
0.1mol/L KCL
30
1.414×10
-2
4. Quality Warranty
1. The quality assurance of the electrode is the period for stored. It is about one year.
2. During  the  warranty  period,  if  you  find  some  problems  caused  by  factories  and  not 
working, the manufacturer should be responsible for the repair or return.