User`s guide

Dexcom G4 PLATINUM User’s Guide | 97
the alert (to change your snooze settings see
Chapter 9, Section 9.2, Advanced Alerts).
9.1.2.3 LOW GLUCOSE ALARM
TheDexcomG4PLATINUMSystemalsohasaxedlowalarmat55
mg/dL. This is different than your low glucose alert. You cannot change
or turn off this alarm or its re-alarm settings.
The receiver displays the low glucose
alarm screen.
Re-alarm: The receiver automatically
alerts again 30 minutes after you press the
SELECT button to clear it, if your sensor
glucose readings are still at or below
55 mg/dL.
Your receiver does not alert if you have a
sensor glucose reading outside your target range and you calibrated
in the last 5 minutes.
The receiver alerts if your sensor glucose reading stays outside your
targetrangeafterveminutes.
9.1.3 GETTING TO THE ALERTS MENU
Low glucose alarm
Low glucose alert setting