User`s guide
94 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
9
Default Alert/Alarm Settings (continued from page before)
What will I see
on the receiver
screen?
What does
this mean?
What is
the default
setting?
How will the
receiver notify
me?
Will the
receiver
re-notify
me?
Low Glucose
Alarm
Your most
recent sensor
glucose
reading is at
or below 55
mg/dL.
On Vibrates 4
times and then
vibrates/beeps
4 times every
5 minutes until
conrmed or
your glucose
value goes
above 55
mg/dL.
Yes, every
30 minutes
after each
conrmation
until your
blood
glucose
value
comes back
into range.
Out of Range
Alert
The transmitter
and receiver
are not
communicating
to each other.
Off The alert will
not notify you.
You must
change the
settings to
receive this
alert.
No
Rise/Fall Alert
Single Arrow
Your glucose
is rising/falling
at or above a
rate of
2 mg/dL/min
(at least 30
mg/dL in 15
minutes).
Off The alert will
not notify you.
You must
change the
settings to
receive this
alert.
No
(continued on next page)