User`s guide

186 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
14
Electromagnetic Emissions Specications (continued
from page before)
Emissions Test Compliance Electromagnetic
Environment Guidance
RF Emissions
CISPR 11
Class B The Dexcom G4 PLATINUM
System is suitable for use
in all establishments other
than domestic and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic Emissions
IEC 61000-3-2
Not applicable
Voltage Fluctuations/Flicker
Emissions
IEC 61000-3-3
Not applicable
Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and
the Receiver
The receiver is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the
user of the receiver can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the receiver as
recommended in the next table, according to the maximum output
power of the communications equipment. Portable and mobile RF
equipment include: baby monitors, Bluetooth
®
wireless headsets,
wirelessrouters,microwaveovens,laptopswithinternalwi-adapters,
GSM cell phones, RFID scanners and hand-held security metal detector
often used by security screeners.