User`s guide
Dexcom G4 PLATINUM User’s Guide | 183
System Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Conducted RF
IEC 61000-4-6
(Receiver only)
3 Vrms
150 kHz to 80 MHz
3 Vrms Portable and mobile
RF communications
equipment should
be used no closer
to any part of the
receiver, including
cables, than the
recommended
separation distance
calculated from the
equation applicable
to the frequency of
the transmitter.
Recommended
Separation
Distance
d = 1.2 √P 150 kHz
to 80 MHz
d = 1.2 √P 80 MHz
to 800 MHz
d = 2.3 √P 800
MHz to 2.5 GHz
Where P is the
maximum output
power rating of the
transmitter in watts
(W) according to
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
20 V/m
80 MHz to 2.4 GHz
10 V/m
2.4 GHz to 2.5 GHz
(continued on next page)