Receiver Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
IEC 61000-4-11
IEC 60601-1-11
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
85% U
T
(15% dip in
U
T
) for 5 sec
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
85% U
T
(15% dip in
U
T
) for 5 sec
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Dexcom G4 PLATINUM System is intended for use in the
electromagneticenvironmentspeciedinthenexttable.Thecustomer
or the user of the Dexcom G4 PLATINUM System should ensure that it
is used in such an environment.