Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The receiver (P/N MT22495) is intended for use in the electromagnetic
environmentspeciedinthenexttable.Thecustomerortheuserofthe
receiver should ensure that it is used in such an environment.
Receiver Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be
wood, concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
Not applicable
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
Not applicable
Mains power quality
should be that of a
typical commercial
or hospital
environment.