User`s guide
Dexcom G4 PLATINUM User’s Guide | 181
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The receiver (P/N MT22495) is intended for use in the electromagnetic
environmentspeciedinthenexttable.Thecustomerortheuserofthe
receiver should ensure that it is used in such an environment.
Receiver Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Receiver
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be
wood, concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
Not applicable
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
Not applicable
Mains power quality
should be that of a
typical commercial
or hospital
environment.