User`s guide

Dexcom G4 PLATINUM User’s Guide | 179
Transmitter Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance Level
Electromagnetic
Environment
Guidance
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
IEC 61000-4-11
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
Not applicable
Battery operated
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.