User`s guide
178 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
14
device can withstand exposure to common electrostatic (ESD) and
electromagnetic interference (EMI).
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-01 and P/N 9438-05) is intended for use
intheelectromagneticenvironmentspeciedinthenexttable.The
customer or the user of the transmitter should ensure that it is used in
such an environment.
Transmitter Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be
wood, concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Not applicable
Battery operated
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not applicable
Battery operated
(continued on next page)