User`s guide

178 | Dexcom G4 PLATINUM Continuous Glucose Monitoring System
14
device can withstand exposure to common electrostatic (ESD) and
electromagnetic interference (EMI).
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-01 and P/N 9438-05) is intended for use
intheelectromagneticenvironmentspeciedinthenexttable.The
customer or the user of the transmitter should ensure that it is used in
such an environment.
Transmitter Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be
wood, concrete or
ceramic tile. If oors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Not applicable
Battery operated
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not applicable
Battery operated
(continued on next page)