Dexcom G4 PLATINUM User’s Guide | 137
SELECT button to clear the screen, this symbol shows in the
status area.
If you see this screen, wait at least 1 hour and
then enter 1 more calibration blood glucose
value. Wait 15 minutes. If this error screen still
shows, enter 1 more blood glucose value. Wait
another 15 minutes. If this error screen still
shows, the sensor needs to be replaced.
13.4 SYSTEM GLUCOSE ERROR
The system may tell you that it cannot provide a sensor glucose reading.
When this happens you will see either the glucose reading error symbol
or the wait symbol in the status area. These symbols mean
the receiver does not understand the sensor signal temporarily. These
symbols are related to the sensor only.
Wait for more prompts, and do not enter any
blood glucose values when you see these
symbols. The system will not use a blood
glucose value for calibration when these
symbols show (see Chapter 8, Section 8.3,
Glucose Status Area Symbols).
Often, the system can correct the problem and
continue providing sensor glucose readings.
If it has been at least 3 hours since your last
sensor glucose reading, contact Technical
Support (see Chapter 15, User Assistance).
If you see these system glucose error icons
often, follow these troubleshooting tips before
inserting another sensor:
• Make sure your sensor is not expired.
Wait 1 hour
calibration error screen
No glucose data ( )
No glucose data ( )