User`s guide

188 | Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
14
Electromagnetic Emissions Specications (continued
from page before)
Emissions Test Compliance Electromagnetic Environment
Guidance
RF Emissions
CISPR 11
Class B The Dexcom G4 PLATINUM (Pediatric)
System is suitable for use in all
establishments other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not applicable
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
Not applicable
Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and
the Receiver
The receiver is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the
user of the receiver can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the receiver as
recommended in the next table, according to the maximum output
power of the communications equipment. Portable and mobile RF
equipment include: baby monitors, Bluetooth
®
wireless headsets,
wirelessrouters,microwaveovens,laptopswithinternalwi-adapters,
GSM cell phones, RFID scanners and hand-held security metal detector
often used by security screeners.