Transmitter Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance Level
Electromagnetic
Environment
Guidance
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Not applicable
Battery operated
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
IEC 61000-4-11
< 5% U
T
(> 95% dip
in U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
< 5% U
T
(> 95% dip
in U
T
) for 5 sec
Not applicable
Battery operated
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic elds
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Receiver Product Specications
Part Number MT22608
(continued on next page)