User`s guide
168 | Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
14
glucose level was at or below the alert setting within 15 or 30 minutes
before or after the device alarmed. The False Alert Rate is the percent
of time the device alarmed when the blood glucose level was above the
alert setting within 15 or 30 minutes before or after the device alarmed.
For example, if you set the Low Glucose Alert to 70 mg/dL (3.9 mmol/L)
and your alarm sounds, how often can you expect your blood sugar
to actually be low? If your alarm sounds, you can expect your blood
sugar to be below 70 mg/dL (3.9 mmol/L) approximately 47% of the time
(pediatric study) or 79% of the time (adult study) and not be below 70
mg/dL (3.9 mmol/L) approximately 53% of the time (pediatric study) or
21 % of the time (adult study) within the 15 minute period before or after
your alarm sounds.
Hypoglycemia Detection Rate
The Detection Rate shows how often the device recognizes and alerts
you to an episode of hypoglycemia or how often it misses such an event.
The Hypoglycemia Detection Rate is the percent of time the blood
glucose level was at or below the alert setting and device alarmed within
15 or 30 minutes before or after the blood glucose was at or below the
alert settings. The Hypoglycemia Missed Detection Rate is the percent
of time the blood glucose was at or below the alert setting, but the
device did not alarm within 15 or 30 minutes before or after the blood
glucose was at or below the alert setting.
For example, if you set the Low Glucose alert to 70 mg/dL (3.9 mmol/L),
how often will your alarm alert you if your blood glucose goes below
70 mg/dL (3.9 mmol/L)? If your blood sugar goes below 70 mg/dL (3.9
mmol/L), you can expect your alarm to sound 50% of the time (pediatric
study) or 83% of the time (adult study) and not to sound approximately
50% of the time (pediatric study) or 17% (adult study) of time within
the 15 minute period before or after your blood sugar goes below 70
mg/dL (3.9 mmol/L).
The High Glucose Alert
Estimates of how well the adjustable High Glucose Alert performs are