User`s guide
162 | Dexcom G4 PLATINUM (Pediatric) Continuous Glucose Monitoring System
14
Table 3-B. Concurrence of CGM Readings and YSI
Values (Adult Study)
CGM
mg/dL
(mmol/L)
YSI mg/dL (mmol/L)
Percent of matched pairs in each YSI glucose range
for each Sensor glucose range
< 40
(< 2.2)
40-60
(2.2-
3.3)
61-80
(3.4-
4.4)
81-
120
(4.5-
6.7)
121-
160
(6.7-
8.9)
161-
200
(8.9-
11.1)
201-
250
(11.1-
13.9)
251-
300
(13.9-
16.7)
301-
350
(16.7-
19.4)
351-
400
(19.4-
22.2)
> 400
(> 22.2)
Number
of
Paired
CGM-YSI
< 40
(< 2.2)
6% 48% 37% 7% 1% 0% 0% 0% 0% 0% 0% 155
40-60
(2.2-3.3)
4% 49% 36% 11% 1% 0% 0% 0% 0% 0% 0% 512
61-80
(3.4-4.4)
0% 22% 51% 24% 1% 0% 0% 0% 0% 0% 0% 781
81-120
(4.5-6.7)
0% 2% 17% 66% 13% 1% 0% 0% 0% 0% 0% 1706
121-160
(6.7-8.9)
0% 0% 1% 25% 60% 13% 2% 0% 0% 0% 0% 1492
161-200
(8.9-11.1)
0% 0% 0% 2% 28% 53% 16% 2% 0% 0% 0% 1240
201- 250
(11.1- 13.9)
0% 0% 0% 0% 3% 21% 51% 21% 3% 1% 0% 1181
251- 300
(13.9- 16.7)
0% 0% 0% 0% 0% 4% 19% 49% 24% 3% 0% 1018
301- 350
(16.7- 19.4)
0% 0% 0% 0% 0% 0% 3% 28% 51% 16% 1% 775
351- 400
(19.4- 22.2)
0% 0% 0% 0% 0% 0% 3% 10% 43% 38% 7% 447
> 400
(> 22.2)
0% 0% 0% 0% 0% 0% 1% 6% 21% 57% 15% 248