User`s guide
Dexcom G4 PLATINUM (Pediatric) System User’s Guide | 161
16
Table 3-A. Concurrence of CGM Readings and YSI
Values (Pediatric Study)
YSI mg/dL (mmol/L)
Sensor
mg/dL
(mmol/L)
< 40
(2.2)
40-
60
(2.2-
3.3)
61-80
(3.4-
4.4)
81-
120
(4.5-
6.7)
121-
160
(6.7-
8.9)
161-
200
(8.9-
11.1)
201-
250
(11.1-
13.9)
251-
300
(13.9-
16.7)
301-
350
(16.7-
19.4)
351-
400
(19.4-
22.2)
> 400
(> 22.2)
Total
< 40
(2.2)
0% 0% 0% 54% 31% 15% 0% 0% 0% 0% 0% 13
40-60
(2.2-3.3)
0% 21% 58% 16% 5% 0% 0% 0% 0% 0% 0% 19
61-80
(3.4-4.4)
0% 21% 45% 30% 4% 0% 0% 0% 0% 0% 0% 76
81-120
(4.5-6.7)
0% 1% 20% 66% 12% 1% 0% 0% 0% 0% 0% 338
121-160
(6.7-8.9)
0% 0% 1% 36% 54% 7% 1% 0% 0% 0% 0% 511
161-200
(8.9-11.1)
0% 0% 0% 4% 40% 48% 6% 1% 0% 0% 0% 596
201- 250
(11.1- 13.9)
0% 0% 0% 1% 9% 44% 41% 5% 0% 0% 0% 658
251- 300
(13.9- 16.7)
0% 0% 0% 0% 2% 7% 50% 36% 3% 0% 2% 432
301- 350
(16.7- 19.4)
0% 0% 0% 0% 0% 2% 18% 59% 21% 0% 0% 206
351- 400
(19.4- 22.2)
0% 0% 0% 0% 0% 0% 3% 28% 50% 16% 2% 86
> 400
(> 22.2)
0% 0% 0% 0% 0% 0% 1% 14% 41% 36% 7% 70