User`s guide

Dexcom G4 PLATINUM (Pediatric) System User’s Guide | 99
Default Alert/Alarm Settings (continued from page before)
What will I see
on the receiver
screen?
What does
this mean?
What is
the default
setting?
How will the
receiver notify
me?
Will the
receiver
re-notify
me?
Rise/Fall Alert
Single Arrow
Your glucose
is rising/falling
at or above a
rate of
2 mg/dL/min
(at least 30
mg/dL in 15
minutes).
Off The alert will
not notify you.
You must
change the
settings to
receive this
alert.
No
Rise/Fall Alert
Double Arrow
Your glucose
is rising/falling
at or above a
rate of
3 mg/dL/min
(at least 45
mg/dL in 15
minutes).
Off The alert will
not notify you.
You must
change the
settings to
receive this
alert.
No
This table describes the receiver alerts and alarm and explains how the
receiver will notify you in the default setting.
Your receiver may alert you at other times you need to take action, such
as low battery, failed sensor, etc. See Chapter 18, Appendix I, for a
detailed list of these other alerts.