HELPFUL HINTS:
• On the rst alert notication the receiver will vibrate only (no beep).
Regardless of the specic alert prole you chose, if you conrm the initial
vibrate alert, you will not receive an additional audible alert.
• The xed low alarm at 3.1 mmol/L cannot be turned off or adjusted.
• For the soft, normal, attentive and hyporepeat proles the following alert
sequence will occur:
• The rst alert is vibrate only.
• If the alert is not conrmed in ve minutes, the system will
vibrate and beep.
• If the alert is not conrmed in ve more minutes, the system
will vibrate and beep louder and this will continue at the same
volume level every ve minutes until conrmed.
For the hyporepeat prole only:
• If the alert is conrmed and your sensor glucose readings
continue to be at or below 3.1 mmol/L your system will repeat
the alert sequence above in 30 minutes.
To conrm any alert press the SELECT button.