User`s guide

54 | Dexcom G4 Continuous Glucose Monitoring System
1. When you see this calibration prompt it means
it is time to calibrate with a single blood glucose
value. You can clear the prompt by pushing the
SELECT button. The system will re-alert you every
15 minutes until the receiver accepts the blood
glucose value.
2. Take 1 blood glucose measurement with your meter
and enter the blood glucose value into the receiver.
If this screen reappears shortly after you have entered a new blood glucose
value see Chapter 9, Section 9.2, Calibration Troubleshooting.
4
Calibration prompt
4.5 OTHER REASONS YOU MAY NEED TO CALIBRATE
• When your system did not accept the last calibration.
When your blood glucose value is very different from the sensor glucose
reading.
1. When you see this calibration prompt it means it is
time to calibrate with a single blood glucose value.
2. Take 1 blood glucose measurement with your meter
and enter the blood glucose into the receiver. If this
screen reappears shortly after you have entered a
new blood glucose value see Chapter 9, Section
9.2.1, Types of Calibration Prompts.
HELPFUL H INT:
If you see these screens they are indicating calibration errors (see Chapter
9, Section 9.3, Calibration Error Troubleshooting).
15 minute calibration
error screen
1 hour calibration
error screen
Calibration prompt