User`s guide
Dexcom G4 User’s Guide | 119
11.1 PRODUCT SPECIFICATIONS
Sensor
Glucose Range 2.2 – 22.2 mmol/L
Sensor Life Up to 7 days
Calibration Commercially available blood glucose meter
Calibration Range 2.2 – 22.2 mmol/L
Storage
Condition
Temperature: 2° C - 25° C
Humidity: 0% - 95% RH
Sterilization Sterile by radiation
Transmitter
Dimensions (including
sensor pod)
Length: 3.8 cm
Width: 2.3 cm
Thickness: 1.3 cm
Weight (including
sensor pod)
10 g
Power Supply Silver oxide batteries (not replaceable)
Operational Conditions Temperature: 10° C - 42° C
Humidity: 10% - 95% RH
Storage Conditions Temperature: 0° C - 45° C
Humidity: 10% - 95% RH
Operating Altitude -152 to 3657 meters
Limited Warranty 6 months
Moisture
Protection
IP28: temporary submersion
Protection Against
Electrical Shock
Type BF applied part