User`s guide
102 | Dexcom G4 Continuous Glucose Monitoring System
9.2.2 STATUS AREA SYMBOLS DURING CALIBRATION
PROMPTS
1. Clear the prompt from the screen by pressing the SELECT button.
2. Check the status area at the top of the screen. You
will decide what to do based on the symbol shown
in the status area:
a. If the startup calibration symbol is displayed:
• Take 2 more blood glucose values and enter
them into your receiver
b. If the additional startup calibration symbol is
displayed:
• Take 1 more blood glucose value and enter it into your receiver.
c. If the calibration needed symbol is displayed:
• Take 1 more blood glucose value and enter it into your receiver.
9
Prompt in status area
9.3 CALIBRATION ERROR TROUBLESHOOTING
This screen means you recently entered a calibration
blood glucose value and the sensor is having trouble
calibrating. If you clear this screen by pressing the
SELECT button this symbol will be displayed in
the status area.
If you see this screen, wait 15 minutes and then enter
1 more calibration blood glucose value. Wait 15 more
minutes. If this error screen still appears enter 1 more
blood glucose value. Wait another 15 minutes. If this
error screen still appears, the sensor needs to be
replaced.
This screen means you recently entered a calibration
blood glucose value and the sensor is having trouble
calibrating. If you clear this screen by pressing the
SELECT button, this symbol will be displayed
in the status area.
Wait 15 minute
calibration error screen
Wait 1 hour calibration
error screen