User`s guide
Dexcom G4 PLATINUM User’s Guide | 199
What will I see
on the receiver
screen?
Prompt,
alert or
alarm?
What does this
mean?
How will the
receiver notify
me?
(vibrate and/or
beep)
Will the
receiver
re-notify
me?
12 hour
calibration
prompt
The receiver
needs a blood
glucose value
entered to
calibrate.
Prompt screen
only.
Yes,
every 15
minutes.
Calibration
prompt
The receiver
needs a blood
glucose value
entered to
calibrate. Sensor
glucose readings
will not be
displayed at this
time.
1 vibrate then
vibrate/beep
every 5 minutes
until conrmed.
Yes,
every 15
minutes.
Startup
calibration
prompt
The receiver
needs 2 blood
glucose values
entered to
calibrate.
1 vibrate then
vibrate/beep
every 5 minutes
until conrmed.
Yes,
every 15
minutes.
Additional
startup
calibration
prompt
The receiver
needs 1
additional blood
glucose value to
complete startup
calibration.
1 vibrate then
vibrate/beep
every 5 minutes
until conrmed.
Yes,
every 15
minutes.
Receiver Alerts, Alarm and Prompts (continued from page
before)
(continued on next page)