User`s guide
Dexcom G4 PLATINUM Professional System User’s Guide | 85
1. Wash and dry hands, make sure the glucose test strips are not
expired and have been stored properly, and make sure the meter is
properly coded (if required).
2. Take a blood glucose measurement using the meter.
•Carefullyapplythebloodsampletotheteststripfollowingthemeter
or test strip instructions.
3. From any trend graph, press the SELECT button to see the Main
Menu.
4. Use the UP or DOWN button to scroll until
you highlight “Enter BG.”
•“EnterBG”willbethesecondMainMenu
option when you are in the middle of a
sensor session.
5. Press the SELECT button to choose this
option. You will see a screen with a blood
drop and a number in mg/dL units.
a. At all times in real-time and blinded modes,
when the receiver does not have a recent
sensor glucose reading the default is 120
mg/dL.
b. If there has been a sensor glucose reading
in the past 15 minutes in real-time mode,
the Enter BG screen shows the current
sensor glucose reading. Do not use the
current sensor glucose reading for
calibration. Use only blood glucose values from the meter.
6. Use the UP or DOWN button to scroll until you see the correct blood
glucose value, and then press the SELECT button.
Main Menu,
Enter BG highlighted
Enter BG screen,
default is 120 mg/dL