User`s guide
76 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
6
1. Press SELECT during the startup period to turn the on receiver
display and see this symbol.
In both real-time and blinded modes, the countdown symbol llsin
overtimetoshowthatyouaregettingclosertotherstcalibrationtime.
You will not get sensor glucose readings, high/low glucose alerts and
alarm during the countdown.
If you see the out of range symbol at the top of the screen during the
2-hour startup, review the following troubleshooting tips:
•Makesurethereceiverandtransmitterarewithin20feetofeach
other without obstruction. Check in 10 minutes to see if the antenna
symbol shows in the upper left corner of the receiver screen.
•Ifthereceiverandtransmitterarestillnotcommunicating,checkthe
device information screen to make sure the correct transmitter ID is
entered into the receiver (see Chapter 5, Section 5.2, The Settings
Menu).
•IfthecorrecttransmitterIDisinthereceiverandthereceiverand
transmitter are still not communicating, contact Dexcom Technical
Support (see Chapter 15, User Assistance).
At the end of the 2-hour startup period the receiver lets you know it
is time to calibrate the sensor. Chapter 7, Calibrating the Dexcom G4
PLATINUM Professional System, tells you how to calibrate the sensor.
6.9 TAPING THE SENSOR POD
The sensor pod should stay on the skin using its own adhesive. But, if
the patch is peeling up, you can use medical tape (such as Blenderm
™
,
Tegaderm
™
, Smith & Nephew IV3000
®
, 3M
™
tape) for extra support.
If you use tape, only tape over the white adhesive patch on all sides
[0-24]
minutes
[24-48]
minutes
[72-96]
minutes
Ready for
calibration
[48-72]
minutes
Start
Countdownsymbolllsinduringsensorstartup