User`s guide
248 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
18
Receiver Alerts, Alarm and Prompts (continued from page
before)
What will I see
on the receiver
screen?
Prompt,
alert or
alarm?
What does this
mean?
How will the
receiver notify
me?
(vibrate and/or
beep)
Will the
receiver
re-notify
me?
In both real-time
and blinded modes:
30-minute
sensor
expiration
alert
The sensor
session will end
in 30 minutes.
Vibrates 1
time and then
vibrates/beeps
every 5 minutes
(2 repeats max).
No
In both real-time
and blinded modes:
End of
session
sensor
expiration
alert
The sensor
session has
ended.
Vibrates 1
time and then
vibrates/beeps
every 5 minutes
(2 repeats max).
No
In both real-time
and blinded modes:
Sensor
failed alert
The sensor is not
working properly.
Vibrates 1
time and then
vibrates/beeps
every 5 minutes
(2 repeats max).
Yes, 2
re-alerts
in the
next 10
minutes
for 30
minutes.
(continued on next page)