User`s guide

Dexcom G4 PLATINUM Professional System User’s Guide | 243
Receiver Alerts, Alarm and Prompts (continued from page
before)
What will I see
on the receiver
screen?
Prompt,
alert or
alarm?
What does this
mean?
How will the
receiver notify
me?
(vibrate and/or
beep)
Will the
receiver
re-notify
me?
In both real-time
and blinded modes:
Wait 1
hour
calibration
error alert
The sensor
cannot calibrate.
Wait a minimum
of 1 hour then
enter 1 more
blood glucose
value for
calibration. If no
sensor glucose
readings appear
on the receiver,
the sensor needs
to be replaced.
Vibrates 1
time and then
vibrates/beeps
every 5 minutes
until conrmed.
No
In both real-time
and blinded modes:
12 hour
calibration
prompt
The receiver
needs a blood
glucose value
entered to
calibrate.
Prompt screen
only.
Yes,
every 15
minutes.