User`s guide
242 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
18
Receiver Alerts, Alarm and Prompts (continued from page
before)
What will I see
on the receiver
screen?
Prompt,
alert or
alarm?
What does this
mean?
How will the
receiver notify
me?
(vibrate and/or
beep)
Will the
receiver
re-notify
me?
In both real-time
and blinded modes:
Wait 15
minutes
calibration
error alert
The sensor
cannot calibrate.
Wait 15 minutes
then enter 1 more
blood glucose
value. Wait 15
more minutes. If
error screen still
appears enter
1 more blood
glucose value.
Wait 15 minutes.
If no sensor
glucose readings
appear on the
receiver, the
sensor needs to
be replaced.
Vibrates 1
time and then
vibrates/beeps
every 5 minutes
until conî‚¿rmed.
No
(continued on next page)