User`s guide
214 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Minimum Recommended Distance Between Transmitter
and Receiver
Rated
maximum
Output Power
of Transmitter
(W)
Separation Distance
According to Frequency of Transmitter (m)
150 kHz to 80
MHz
d = 1.2 P
1/2
80 MHz to 800
MHz
d = 1.2 P
1/2
800 MHz to 2.5
GHz
d = 2.3 P
1/2
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance (d) in feet can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacture.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reection
from structures, objects and people.
USB Charging/Download Cable* Specications
Dexcom P/N MT20655
Input/Output 5V DC, 1A
Type USB A to USB micro B
Length 3 feet
* The power supply/charger can be connected to the USB
charging/download cable for charging using an AC power outlet.