14
a.Fieldstrengthsfromxedtransmitters,suchasbasestationsfor
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic
environmentduetoxedRFtransmitters,anelectromagneticsite
surveyshouldbeconsidered.Ifthemeasuredeldstrengthinthe
location in which the Dexcom G4 PLATINUM Professional System is
used exceeds the applicable RF compliance level above, the Dexcom
G4 PLATINUM Professional System should be observed to verify
normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the
Dexcom G4 PLATINUM Professional System.
b.Overthefrequencyrange150kHzto80MHz,eldstrengthsshould
be less than 3 V/m.
Guidance and Manufacturer’s Declaration –
Electromagnetic Emissions
The Dexcom G4 PLATINUM Professional System is intended for use
intheelectromagneticenvironmentspeciedinthenexttable.The
customer or the user of the Dexcom G4 PLATINUM Professional
System should ensure that it is used in such an environment.
Electromagnetic Emissions Specications
Emissions Test Compliance Electromagnetic Environment
Guidance
RF emissions
CISPR 11
Group 1 The Dexcom G4 PLATINUM
Professional System uses
RF energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
(continued on next page)