User`s guide
210 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
System Electromagnetic Immunity Specications
Immunity Test IEC 60601
Test Level
Receiver
Compliance
Level
Electromagnetic
Environment
Guidance
Conducted RF
IEC 61000- 4-6
(Receiver only)
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 Vrms
10 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the receiver,
including cables, than
the recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
Separation Distance
d=1.2√P150kHzto
80 MHz
d=1.2√P80MHzto
800 MHz
d=2.3√P800MHzto
2.5 GHz
(continued on next page)