User`s guide
Dexcom G4 PLATINUM Professional System User’s Guide | 209
Receiver Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601
Test Level
Receiver
Compliance
Level
Electromagnetic
Environment
Guidance
IEC 61000-4-11
IEC 60601-1-11
70%U
T
(30%dipinU
T
)
for 25 cycles
85%U
T
(15%dipinU
T
)
for 5 sec
<5%U
T
(>95%dipinU
T
)
for 5 sec
70%U
T
(30%dipinU
T
)
for 25 cycles
85%U
T
(15%dipinU
T
)
for 5 sec
<5%U
T
(>95%dipinU
T
)
for 5 sec
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magnetic
eldsshould
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Dexcom G4 PLATINUM Professional System is intended for use
intheelectromagneticenvironmentspeciedinthenexttable.The
customer or the user of the Dexcom G4 PLATINUM Professional
System should ensure that it is used in such an environment.