User`s guide

208 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Receiver Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601
Test Level
Receiver
Compliance
Level
Electromagnetic
Environment
Guidance
Electrical Fast
Transient
/burst
IEC 61000-4-4
±2kVfor
power supply
lines
±1kVfor
input/output
lines
±2kVfor
power supply
lines
Not applicable
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge
IEC 61000-4-5
±1kVline(s)
to line(s)
±2kVline(s)
to earth
±1kVline(s)
to line(s)
Not applicable
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
<5%U
T
(>95%dipinU
T
)
for 0.5 cycle
40%U
T
(60%dipinU
T
)
for 5 cycles
<5%U
T
(>95%dip
in U
T
) for 0.5
cycle
40%U
T
(60%dipinU
T
)
for 5 cycles
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
(continued on next page)