User`s guide
208 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Receiver Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601
Test Level
Receiver
Compliance
Level
Electromagnetic
Environment
Guidance
Electrical Fast
Transient
/burst
IEC 61000-4-4
±2kVfor
power supply
lines
±1kVfor
input/output
lines
±2kVfor
power supply
lines
Not applicable
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Surge
IEC 61000-4-5
±1kVline(s)
to line(s)
±2kVline(s)
to earth
±1kVline(s)
to line(s)
Not applicable
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
Voltage
Dips, Short
Interruptions
and Voltage
Variations on
Power Supply
Input Lines
<5%U
T
(>95%dipinU
T
)
for 0.5 cycle
40%U
T
(60%dipinU
T
)
for 5 cycles
<5%U
T
(>95%dip
in U
T
) for 0.5
cycle
40%U
T
(60%dipinU
T
)
for 5 cycles
Mains power
quality should be
that of a typical
commercial
or hospital
environment.
(continued on next page)