User`s guide

206 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
Transmitter Electromagnetic Immunity Specications
(continued from page before)
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance
Level
Electromagnetic
Environment
Guidance
Power
Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency
magneticelds
should be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Receiver Product Specications
Part Number MT20649
Reading Frequency Every 5 minutes
Dimensions
Length: 4.0 inches
Width: 1.8 inches
Thickness: 0.5 inches
Weight 2.4 ounces
Receiver Input 5V DC, 1A
Power Supply MT21255
Communication Range 20 feet
Memory Storage
30 days of glucose data,
7 days of tech support data
Re-Chargeable Battery Use 3 days
(continued on next page)