User`s guide
204 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
14
The Dexcom G4 PLATINUM Professional Continuous Glucose
Monitoring System is safe for use on U.S. commercial airlines. The
Dexcom G4 PLATINUM Transmitter is an M-PED with emission levels
that meet RTCA/DO160, Section 21, Category M. Per FAA Advisory,
Circular #91-21, 1B, dated 8/25/06, any M-PED that meets this standard
in all modes may be used onboard the aircraft without any further
testing by the operator. This device can withstand exposure to common
electrostatic (ESD) and electromagnetic interference (EMI).
Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The transmitter (P/N 9438-01) is intended for use in the electromagnetic
environmentspeciedinthenexttable.Thecustomerortheuserofthe
transmitter should ensure that it is used in such an environment.
Transmitter Electromagnetic Immunity Specications
Immunity Test IEC 60601 Test
Level
Transmitter
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
±6kVContact
±8kVAir
±6kV
Contact
±8kVAir
Floors should be
wood, concrete
or ceramic
tile.Ifoors
are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical Fast
Transient
/burst
±2kVfor
power supply
lines
Not applicable
(continued on next page)