202 | Dexcom G4 PLATINUM Professional Continuous Glucose Monitoring System
PercentDifferenceshowsthathalfofthetimetheSystemreads-1.4%
or less than the SMBG values and the Median Absolute Percent
DifferenceshowsthathalfofthetimetheSystemreadsabout11.0%or
less different than SMBG values (Table 12).
Adverse Events
No serious adverse events or device-related serious adverse events
occurred during the study. Mild skin irritation, such as erythema or
edema, occurred in low frequency around adhesive area. No infection,
bruising, or bleeding occurred at the sensor needle insertion area or the
adhesive area.
14.2 PRODUCT SPECIFICATIONS
Sensor Product Specications
Glucose Range 40 - 400 mg/dL
Sensor Life Up to 7 days
Calibration Commercially available blood glucose meter
Calibration Range 40 - 400 mg/dL
Storage
Condition
Temperature: 36° F - 77° F
Humidity:15%-85%RH
Sterilization Sterile by radiation
Transmitter Product Specications
Part Number 9438-01
Dimensions
(including
sensor pod)
Length: 1.5 inches
Width: 0.9 inches
Thickness: 0.5 inches
Weight (including
sensor pod)
0.4 ounces