User`s guide

Dexcom G4 PLATINUM Professional System User’s Guide | 109
sensor glucose readings drop below the high
glucose alert level. You can have the receiver
re-alert after clearing the alert (to change the
snooze settings see Chapter 9, Section 9.2,
Advanced Alerts).
9.1.2.2 LOW GLUCOSE ALERT IN REAL-TIME
MODE
When the sensor glucose readings are at or
below the low alert level, this screen shows the
low glucose alert level. The receiver vibrates
and/orbeepsdependingontheprolesetting.
This level shows as a red line on the trend
graph.
The receiver continues to alert until you press
the SELECT button to clear the alert or until
the sensor glucose readings rise above the
low glucose alert level. You can have the
receiver re-alert after clearing the alert (to
change the snooze settings see Chapter 9,
Section 9.2, Advanced Alerts).
9.1.2.3 LOW GLUCOSE ALARM IN REAL-TIME
MODE
TheDexcomG4PLATINUMProfessionalSystemalsohasaxedlow
alarm at 55 mg/dL. This is different than the low glucose alert. You
cannot change or turn off this alarm or its re-alarm settings.
High glucose alert setting
Low glucose alert setting
Low glucose alert set
at 80 mg/dL