User's Manual
75
The receiver shows this prompt screen every 15 minutes until the receiver accepts the blood glucose
value.
13.2.2 WHAT TO DO FOR CALIBRATION PROMPTS
1. When you see a calibration prompt, press the
SELECT
button to clear the prompt.
2. Check the status area at the top of the screen.
a. If the startup calibration symbol shows, take 2 more blood glucose values and enter them into
your receiver.
b. If the additional startup calibration symbol shows, take 1 more blood glucose value and enter it
into your receiver.
c. If the calibration needed symbol shows, take 1 more blood glucose value and enter it into your
receiver.
13.3 CALIBRATION ERROR TROUBLESHOOTING
This screen means you recently entered a calibration blood glucose value, and the sensor is having
trouble calibrating. If you press the
SELECT
button to clear this screen, this symbol
shows in the
status area.