User's Manual

115
Out of Range
alert
The transmitter
and receiver are
not
communicating
and you will not
receive sensor
glucose readings.
Vibrates 1 time and then
vibrates/beeps every 5
minutes until the receiver
and transmitter are back in
range.
No unless you
have turned on
the out of range
alert.
Unknown
sensor prompt
The sensor is
sending sensor
glucose readings
that the receiver
does not
understand. You
will not receive
sensor glucose
readings.
Symbol in status area only.
N/A
Wait prompt
The receiver has
detected a
potential problem
with the sensor
signal. You
should wait about
30 minutes for
more prompts.
Do not enter any
blood glucose
values during this
time. You will not
receive sensor
glucose readings.
Symbol in status area only.
N/A
Wait 15
minutes
calibration
error alert
The sensor
cannot calibrate.
Wait 15 minutes
then enter 1
more blood
glucose value.
Wait 15 more
minutes. If error
screen still
appears enter 1
more blood
glucose value.
Wait 15 minutes.
If no sensor
Vibrates 1 time and then
vibrates/beeps every 5
minutes until confirmed.
No