User's Manual

105
Interruptions and
Voltage Variations on
Power Supply Input
Lines
(> 95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
(> 95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
should be that of a
typical commercial or
hospital environment.
IEC 61000-4-11
IEC 60601-1-11
70% U
T
(30% dip in U
T
)
for 25 cycles
85% U
T
(15% dip in U
T
)
for 5 sec
< 5% U
T
(> 95% dip in U
T
) for 5 sec
70% U
T
(30% dip in U
T
) for 25
cycles
85% U
T
(15% dip in U
T
) for 5 sec
< 5% U
T
(> 95% dip in U
T
) for 5 sec
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Dexcom G4 PLATINUM System is intended for use in the electromagnetic environment specified in the
next table. The customer or the user of the Dexcom G4 PLATINUM System should ensure that it is used in
such an environment.
System Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test Level Receiver
Compliance Level
Electromagnetic
Environment Guidance
Conducted RF
IEC 61000- 4-6
(Receiver only)
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
20 V/m
80 MHz to 2.4 GHz
10 V/m
2.4 GHz to 2.5 GHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the receiver,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d = 1.2 P 150 kHz to 80
MHz