User's Manual
103
relative humidity should
be at least 30%
Electrical Fast
Transient /burst
± 2 kV for power supply
lines
Not applicable
IEC 61000-4-4
± 1 kV for input/output
lines
Battery operated
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Not applicable
Battery operated
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
< 5% U
T
(> 95% dip in U
T
) for 5
sec
Not applicable
Battery operated
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
3 A/m
3 A/m Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Receiver Product Specifications
Part Number
MT22495
Reading Frequency
Every 5 minutes
Dimensions
Length: 4.0 inches
Width: 1.8 inches
Thickness:
0.5 inches
Weight
2.4 ounces
Receiver Input
5V DC, 1A