User's Manual

G4-Global System User’s Guide | 65
5.3.3 LOW GLUCOSE ALARM
The G4-Global System also has an automatic Low Glucose ALARM set at
55 mg/dL. This ALARM is a feature in addition to your personal Low and High
glucose Alerts. You cannot change or turn off this ALARM or its Re-ALARM
settings.
When your glucose reading is at or below 55 mg/dL the Receiver will notify you
with 3 ALARMS (each 5 minutes apart) that beep/vibrate as follows:
- 1st ALARM: Vibrate
- 2nd ALARM: Vibrate, then a series of rapid descending tones followed by
3 constant tones. Series is repeated 3 times.
- 3rd ALARM: Vibrate, then a series of rapid descending tones followed by
3 constant tones. Series is repeated 3 times. (louder)
The Receiver will vibrate and beep every 5 minutes after the 3rd ALARM if your
readings are still at or below 55 mg/dL, until you press any Receiver button to
clear it.
For Low Glucose ALARMs the Receiver will display
the screen to the right.
Re-ALARM: The Receiver will automatically notify
you again in 30 minutes after you press any
Receiver button to clear it, if your glucose readings
are still at or below 55 mg/dL.
Low Glucose ALARM
5.3.4 RISE AND FALL GLUCOSE RATE ALERTS
You can activate these Alerts (see Chapter 8, Section
8.4), to have the Receiver tell you if your glucose
readings are rising or falling fast.
You can set the Rise Rate (how fast you are going
up) to one of two different Rise Rate levels, or choose
to turn this Alert off. Depending on your selection,
you will be noti ed if your glucose level is rising
2 mg/dL or more each minute, or 3 mg/dL or more
each minute.