User's Manual

52 | Continuous Glucose Monitoring System
1. If you see the Single Blood Drop Prompt shown
to the right, clear this display by pressing any
Receiver button. The Single Blood Drop Prompt
will not go away until you press a button.
2. Take a ngerstick reading at this time. Enter this reading in the Receiver to
update calibration.
3. This screen will appear every 15 minutes until you enter a new ngerstick
reading that is accepted for calibration.
4. If this screen reappears shortly after you have entered a new ngerstick
reading, review Chapter 5, Section 5.4.
4.5 RE-CALIBRATION
When you enter a ngerstick reading for 12-Hour
Update Calibration (Section 4.3), the System checks
how well it is functioning compared to the blood
glucose meter. During each Calibration Update
attempt, the System may tell you that another
ngerstick reading is needed. This Re-calibration can
also be prompted during Startup Calibration. The
Single Blood Drop Prompt will appear (shown to the
right).
You can clear this Screen by pressing any Receiver
button. The display will return to the Trend Graph. A
Blood Drop Symbol
will appear on the top right of
the Trend Graph screen.
This Blood Drop Symbol
tells you that the System
needs at least one more ngerstick reading. You will
not see glucose readings or trends, or get Alerts/Low
Glucose ALARM when the Blood Drop Symbol
is
still displayed.
4
Single Blood Drop Prompt
Single Blood Drop Prompt
Awaiting Fingerstick
Reading (Blood Drop)
Symbol
No glucose data