User's Manual
50 | Continuous Glucose Monitoring System
HELPFUL HINTS:
• Make sure a glucose reading is displayed at the top of the Trend Graph,
or a Blood Drop Symbol
is displayed in the top right corner of the Trend
Graph, before calibration.
• Only fi ngerstick readings between 40-400 mg/dL can be used for
calibration. If 1 or more of your readings entered was outside of this range,
the Receiver will not calibrate. You will have to wait to until your blood
glucose is in this range to calibrate.
• Always make sure the Antenna Symbol
is displayed in the upper left
corner of the Trend Graph before you enter fi ngerstick readings for
calibration.
• You should always use the same meter you routinely use to measure
your blood glucose to calibrate. Do not switch your meter in the middle of
Sensor session.
4
4.3 STARTUP CALIBRATION
One (1) hour after you start the Sensor session (see
Chapter 3, Section 3.7) the Receiver will tell you that
you need to calibrate. The Receiver will show the
Double Blood Drop Prompt. If your Alerts are set to
the factory default settings, the Receiver will also
vibrate and beep to tell you it is ready for calibration.
1. You can also set your Receiver to vibrate only (see
Chapter 8, Section 8.5).
2. When you see this screen, press any Receiver button to clear the display.
3. Take 2 fi ngerstick readings with your meter and enter them into the Receiver
(see Section 4.2).
NOTE: Your fi ngerstick readings must be between 40 mg/dL and 400
mg/dL for calibration. The Receiver cannot use fi ngerstick readings outside
of this range for calibration.
Double Blood Drop Prompt